| Warranty Period | 5-year (Refer to Instruction Manuals) |
|---|
Make sure necessary tests will be carried out on your end equipment with the power supply installed in accordance with any required EMC/EMI regulations.
Features of UMPS60F48-E
• For medical electric equipment• Medical Isolation Grade 2MOPP
• 4kV isolation
• Suitable for BF application
• Low leakage current
• Economical design
• Class ll
Ordering Information and Options
UMP(1) S(2) 60(3) F(4) □(5) - □
(1) Series name(2) Single output
(3) Output wattage
(4) Universal input
(5) Output voltage
Recommended EMI Filters
Specification
| Contents | Specifications | |
|---|---|---|
| Input Voltage | AC85 ~ 264V 1Φ | |
| Output | Wattage | 60W |
| Voltage/Current | 48V 1.25A | |
| Adjustable Voltage Range | - | |
| Protection Circuit And Others | Overcurrent Protection | Works over 105% of rating and recovers automatically |
| Overvoltage Protection | 55.20 to 67.20V | |
| Remote ON/OFF | - | |
| Remote Sensing | - | |
| Others | - | |
| Operating Temperature/Humidity | -20 to +70℃ (Required to Derating) , 20 - 90%RH (Non condensing) | |
| Safety And Noise Regulations | Agency Approvals (At only AC input) | ANSI/AAMI ES60601-1, EN60601-1 3rd, C-UL (equivalent to CAN/CSA-C22.2 No.60601-1), UL62368-1, C-UL (equivalent to CAN/CSA-C22.2 No.62368-1), EN62368-1, complies with IEC60335-1 |
| Conducted Noise | Complies with FCC Part 15 -B, FCC Part 18 -B, CISPR11-B, CISPR32-B, EN55011-B, EN55032-B | |
| High Attenuation | Complies with IEC61000-3-2 (Class A) (Not built-in to active filter) * | |
| Others | Dimensions (WxHxD) | 55.9x30.5x81.3mm |
| Weight | 200g max | |
| Cooling Method | Convection | |
| Warranty Period | 5-year (Refer to Instruction Manuals) | |
- * For further information and how to use, please see in Cosel product catalog.
Documents Download
| Technical Data | ||
|---|---|---|
| CAD Data | DXF | |
| 3D CAD Data | STEP | |
| IGES | ||
- * For another documents aren't listed, please feel free to contact to a Cosel representative.